Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Samsung GU60A system- Digital X-ray Imaging Systems are intended for Recalled by NeuroLogica Corporation Due to When performing the Stitching function on GU60A system,...

Date: May 24, 2017
Company: NeuroLogica Corporation
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact NeuroLogica Corporation directly.

Affected Products

Samsung GU60A system- Digital X-ray Imaging Systems are intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications

Quantity: 87 units

Why Was This Recalled?

When performing the Stitching function on GU60A system, unintended area is exposed to radiation

Where Was This Sold?

This product was distributed nationwide across all 50 states.