Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Samsung GM85 Mobile Digital X-ray Imaging System Product Usage: The Recalled by NeuroLogica Corporation Due to Potential issue related to the operation of the...

Date: May 6, 2019
Company: NeuroLogica Corporation
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact NeuroLogica Corporation directly.

Affected Products

Samsung GM85 Mobile Digital X-ray Imaging System Product Usage: The GM85 Digital X-ray imaging System is intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications

Quantity: 288 systems

Why Was This Recalled?

Potential issue related to the operation of the safety latch within the column that supports the arm due to a possibility of potential bodily harm due to a fall of the arm in an unlikely event that the column wire breaks.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About NeuroLogica Corporation

NeuroLogica Corporation has 13 total recalls tracked by RecallDetector.

Related Recalls