Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Samsung GM60A-40S Intended for use in generating radiographic images of Recalled by NeuroLogica Corporation Due to When the system is driving over a non-conductive...

Date: September 14, 2015
Company: NeuroLogica Corporation
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact NeuroLogica Corporation directly.

Affected Products

Samsung GM60A-40S Intended for use in generating radiographic images of human anatomy.

Quantity: 7 devices

Why Was This Recalled?

When the system is driving over a non-conductive floor and then over a conductive, grounded part, proper control of the drive behavior may not be possible. The system can be stopped by releasing the drive handle or by pressing the emergency switch. Proper behavior is re-enabled when the system is rebooted.

Where Was This Sold?

This product was distributed to 4 states: MA, MO, NC, OH