Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
Cannulated graduated drill bit 1.9 mm diameter Recalled by NEOSTEO Due to Traces of corrosion on the laser marking was...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact NEOSTEO directly.
Affected Products
Cannulated graduated drill bit 1.9 mm diameter, with AO QC system, REF: INS-114 for use in orthopedic surgery.
Quantity: 25
Why Was This Recalled?
Traces of corrosion on the laser marking was detected on cannulated graduated drill bit, which could lead to the emission of wear / corrosion particles in the operated area, potentially leading to the following risks: Inflammation, pain, allergy, adverse tissue reaction. Device description: The drill bit is used to create the hole required for the introduction of the screw, it is suitable for Neosteo compression screws for cortical and spongy bone. Drilling depth can be read on the instrument. The shaft is coupled with the ¿4.5mm surgical motor using a AO ¿4.5mm male drive connexion. The countersink bit is coupled to a ¿1.0mm k-wire to ensure guiding while drilling. The instrument is packaged in a plastic sheath sealed by welding.
Where Was This Sold?
This product was distributed nationwide across all 50 states.
About NEOSTEO
NEOSTEO has 1 total recall tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report