Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

EasyOne Pro LAB VOS Respiratory Analysis System Pulmonary Function Testing Recalled by NDDd Medical Technologies, Inc. Due to The CO/CO2 sensor cable harness may have lower...

Date: August 15, 2017
Company: NDDd Medical Technologies, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact NDDd Medical Technologies, Inc. directly.

Affected Products

EasyOne Pro LAB VOS Respiratory Analysis System Pulmonary Function Testing Device, Model 3100-1

Quantity: 37 total devices

Why Was This Recalled?

The CO/CO2 sensor cable harness may have lower electromagnetic shielding, which might have resulted in a non-compliance with the IEC 60601-1-2 standard. As a consequence of a lower electromagnetic shielding provided by the cable harness, the CO/CO2 sensor could have been influenced by high frequencies emitted by very close devices.

Where Was This Sold?

This product was distributed to 5 states: KY, MO, NY, OH, PA

Affected (5 states)Not affected

About NDDd Medical Technologies, Inc.

NDDd Medical Technologies, Inc. has 2 total recalls tracked by RecallDetector.