Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

natus newborn care Olympic Brainz Monitor Recalled by Natus Neurology DBA Excel Tech., Ltd. (XLTEK) Due to The OBM system functional test (impedance and noise)...

Date: March 5, 2020
Company: Natus Neurology DBA Excel Tech., Ltd. (XLTEK)
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Natus Neurology DBA Excel Tech., Ltd. (XLTEK) directly.

Affected Products

natus newborn care Olympic Brainz Monitor, REF OBM00002. electroencephalograph

Quantity: 40 units

Why Was This Recalled?

The OBM system functional test (impedance and noise) was not carried out on the affected items prior to release to the customer. The manufacturer cannot attest to the system functionality of the affected items.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Natus Neurology DBA Excel Tech., Ltd. (XLTEK)

Natus Neurology DBA Excel Tech., Ltd. (XLTEK) has 7 total recalls tracked by RecallDetector.