Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Medix NatalCare ST-LX Intensive Care Incubator. Neonatal incubator. Recalled by Natus Medical Incorporated Due to NatalCare LX Incubators were manufactured using a defective...

Date: January 8, 2016
Company: Natus Medical Incorporated
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Natus Medical Incorporated directly.

Affected Products

Medix NatalCare ST-LX Intensive Care Incubator. Neonatal incubator.

Quantity: 75 units (65 units in the US and 10 units in Pakistan) updated in final status report 36 customers with 96 units.

Why Was This Recalled?

NatalCare LX Incubators were manufactured using a defective oval port hinge assembly with the result that the screws holding the port in place can become loose and possibly result in the port detaching from the incubator.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Natus Medical Incorporated

Natus Medical Incorporated has 18 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report