Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Natus Warmette Blanket Cabinet and Warmette Dual Cabinet The Natus Recalled by Natus Manufacturing Limited Due to The fuse blows during initial power up or...

Date: October 15, 2016
Company: Natus Manufacturing Limited
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Natus Manufacturing Limited directly.

Affected Products

Natus Warmette Blanket Cabinet and Warmette Dual Cabinet The Natus Warmette Blanket Cabinet is designed to store and warm blankets and hospital linens; while, the Natus Warmette Dual Cabinet is designed to store and warm blankets and hospital linens in one compartment and fluids (irrigation and/or injection) in the other compartment within the storage times and warming temperatures recommended by the manufacturers of those items

Quantity: 31

Why Was This Recalled?

The fuse blows during initial power up or after a period of time. This results in a complete loss of power to the Cabinet and in some cases has resulted in the fuse cover for the power inlet melting slightly.

Where Was This Sold?

This product was distributed to 12 states: CA, FL, IL, IN, MA, MS, NJ, NY, NC, TX, VA, WA

Affected (12 states)Not affected

About Natus Manufacturing Limited

Natus Manufacturing Limited has 2 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report