Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Myoscience brand iovera Smart Tip Recalled by Myoscience Inc Due to The outer box label of the product was...

Date: June 21, 2013
Company: Myoscience Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Myoscience Inc directly.

Affected Products

Myoscience brand iovera Smart Tip, Myoscience Cryo-Touch IV, a Cryogenic Surgical device, REF Catalog Number: STA0311-10. Product Usage: The Myoscience Cryo-Touch IV (a.k.a. iovera) is used to destroy tissue during surgical procedures by applying freezing cold. It can also be used to produce lesions in peripheral nervous tissue by the application of cold to the selected site for the blocking of pain. The Cryo-Touch IV (a.k.a. iovera) is not indicated for treatment of central nervous system tissue.

Quantity: 3 units

Why Was This Recalled?

The outer box label of the product was labeled with a down revision label which did not include the US labeling requirements, specifically the symbol descriptors and also the "Caution: Federal Law (USA) restricts this device to sale by or on the order of a physician." This product was previously labeled for EU distribution only.

Where Was This Sold?

This product was distributed to 2 states: CA, ID

Affected (2 states)Not affected

About Myoscience Inc

Myoscience Inc has 2 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report