Class I — Serious Health Hazard

Serious health hazard — there is a reasonable probability that use of or exposure to this product will cause serious adverse health consequences or death.

Multi Joystick R-net. Electrical wheelchair component. Recalled by mo-Vis BVBA Due to Devices with firmware versions 2.3 and lower may...

Date: July 10, 2025
Company: mo-Vis BVBA
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact mo-Vis BVBA directly.

Affected Products

Multi Joystick R-net. Electrical wheelchair component.

Quantity: 3 units

Why Was This Recalled?

Devices with firmware versions 2.3 and lower may experience timing variations between software tasks, which could result in skipping the necessary neutral check and the wheelchair producing unintended movement upon power-up when the joystick is out of neutral.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About mo-Vis BVBA

mo-Vis BVBA has 7 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report