Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

AIRO Mobile CT System Recalled by Mobius Imaging, LLC Due to If the tilt-drive motor has an intermittent connection...

Date: September 26, 2017
Company: Mobius Imaging, LLC
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Mobius Imaging, LLC directly.

Affected Products

AIRO Mobile CT System, Model Number: MobiCT-32 Product Usage: The AIRO¿ is intended to be used for X-ray computed tomography applications for anatomy that can be imaged in the 107cm aperture excluding pediatric patients and patients weighing over 400 lbs. (182 kg).

Quantity: 105

Why Was This Recalled?

If the tilt-drive motor has an intermittent connection or broken encoder or sensor wire there is a risk of unintended or unpredictable device motion while the CT system is homing the axis at startup or preparing for a scan.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Mobius Imaging, LLC

Mobius Imaging, LLC has 7 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report