Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Moberg Component Neuromonitoring System (CNS Monitor)-Intended as a short-term neurological Recalled by Moberg Research, Inc. Due to Battery leakage can cause corrosion to the metal...

Date: February 20, 2023
Company: Moberg Research, Inc.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Moberg Research, Inc. directly.

Affected Products

Moberg Component Neuromonitoring System (CNS Monitor)-Intended as a short-term neurological monitoring system that collects, displays, and stores multiple physiological measurements Model: CNS-350

Quantity: 89 units

Why Was This Recalled?

Battery leakage can cause corrosion to the metal enclosure near the electrical interface. Prolonged contact to corrosive materials could cause injury to skin. DOC

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Moberg Research, Inc.

Moberg Research, Inc. has 1 total recall tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report