Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

216 Wilson Frame Pads Recalled by Mizuho OSI Due to Mizuho has discovered that the 5319-37 Wilson Frame...

Date: March 28, 2017
Company: Mizuho OSI
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Mizuho OSI directly.

Affected Products

216 Wilson Frame Pads, Model # 5319-37; Used with Radiolucent Wilson Frames, model numbers 5319G and 5321G. General and Plastic Surgery: These pads are used as a component of the Wilson Frames, which support the torso of a prone patient during spine surgery. The pads are supplied with the frames and as replacements.

Quantity: 216 sets

Why Was This Recalled?

Mizuho has discovered that the 5319-37 Wilson Frame Pads have a design deficiency in the construction of their mounting sleeves that may allow the patient to contact the underlying carbon fiber structure of the Wilson frame and could lead to a pressure injury.

Where Was This Sold?

United States, Canada, Japan, Australia, New Zealand, Great Britain, Sweden, Spain, Germany, Switzerland, Kenya, Saudi Arabia, Hong Kong, South Korea.

About Mizuho OSI

Mizuho OSI has 6 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report