Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

TD60 display Telemetry Transmitter Recalled by Mindray DS USA, Inc. dba Mindray North America Due to Mindray has become aware that the TD60 Display...

Date: September 8, 2016
Company: Mindray DS USA, Inc. dba Mindray North America
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Mindray DS USA, Inc. dba Mindray North America directly.

Affected Products

TD60 display Telemetry Transmitter, SHENZHEN Mindray Bio-Medical Electronics Co., LTD, Mindray Building, Keji 12th Road South, Hi-Tech Industrial Partk, Nanshan, Shenzhen, 518057 P.R. China

Quantity: 1,278 units

Why Was This Recalled?

Mindray has become aware that the TD60 Display Telemetry Transmitter used with the BeneVision Central Station may crack on the bezel surrounding the display and also the transmitter may not power on or will power on but display an error message.

Where Was This Sold?

US and Canada