Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
Rosebud Vital Signs Monitor with NIBP valve assembly board Recalled by Mindray DS USA, Inc. dba Mindray North America Due to NIBP valve assembly board subject to this recall...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact Mindray DS USA, Inc. dba Mindray North America directly.
Affected Products
Rosebud Vital Signs Monitor with NIBP valve assembly board, p/n 115-017679-00, containing software version 8.1.5 - Product Usage: The monitor is intended for spot-check monitoring physiologic parameters, including Pulse Oximetry (SpO2), Pulse Rate (PR), Non Invasive Blood Pressure (NIBP), and Temperature (TEMP) on adult, pediatric, and neonatal patients in healthcare facilities by clinical physicians or appropriate medical staff under the direction of physicians
Quantity: 9 units
Why Was This Recalled?
NIBP valve assembly board subject to this recall includes enhancements that are not included in the current marketing clearance:1. Anti-motion feature 2. NIBP hose blockage identification function
Where Was This Sold?
This product was distributed nationwide across all 50 states.
About Mindray DS USA, Inc. dba Mindray North America
Mindray DS USA, Inc. dba Mindray North America has 39 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report