Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Panorama Patient Monitoring Network Mindray DS USA The Panorama Patient Recalled by Mindray DS USA, Inc. dba Mindray North America Due to Panorama Central Station including the work station View...

Date: March 9, 2016
Company: Mindray DS USA, Inc. dba Mindray North America
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Mindray DS USA, Inc. dba Mindray North America directly.

Affected Products

Panorama Patient Monitoring Network Mindray DS USA The Panorama Patient Monitoring can view recal time, store, print, graph and trend patient clinical and demographic data. The Panoram Patient Monitoring Network can sent independent alarm limits for data send by the bedside monitor.

Quantity: 1447 units US, 45 units OUS

Why Was This Recalled?

Panorama Central Station including the work station View Station, View Station and the eGateway will revert to the year 2000 when Daylight Savings Time (DST) occurs on March 13, 2016, or under various conditions.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Mindray DS USA, Inc. dba Mindray North America

Mindray DS USA, Inc. dba Mindray North America has 39 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report