Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Isoflurane E_Vap.(Safety Filling adapter): used with the A9 Anesthesia System-Used Recalled by Mindray DS USA, Inc. dba Mindray North America Due to V90 electronic vaporizer anesthetic agent output may be...

Date: July 22, 2024
Company: Mindray DS USA, Inc. dba Mindray North America
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Mindray DS USA, Inc. dba Mindray North America directly.

Affected Products

Isoflurane E_Vap.(Safety Filling adapter): used with the A9 Anesthesia System-Used to provide the vapor of an anesthetic agent with controllable concentration to an A9 anesthesia system. Part Number: Part Number 115-066758-00

Quantity: 24 units

Why Was This Recalled?

V90 electronic vaporizer anesthetic agent output may be higher than the set amount due to a component within the vaporizer causing an incomplete seal in some cases. Should this issue occur, the A9 Anesthesia System will generate alarms associated with abnormal vaporizer output. If the anesthetic concentration is too high, the patient can inhale too much anesthetic agent; the patient may suffer from symptoms such as vertigo and drowsiness after the operation. In severe cases, the patient may suffer from neurological sequelae such as hypomnesia and inattention

Where Was This Sold?

US Distribution to states of: California, Georgia, Kentucky, Mississippi, New Jersey, Puerto Rico, Texas, and West Virginia; and International (Foreign) to: Canada

About Mindray DS USA, Inc. dba Mindray North America

Mindray DS USA, Inc. dba Mindray North America has 39 total recalls tracked by RecallDetector.