Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

MICRO-X ROVER MOBILE X-RAY SYSTEM Recalled by Micro-X Ltd. Due to Software calibration error with product equip with a...

Date: August 31, 2022
Company: Micro-X Ltd.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Micro-X Ltd. directly.

Affected Products

MICRO-X ROVER MOBILE X-RAY SYSTEM, MXU-RV19

Quantity: 42 units

Why Was This Recalled?

Software calibration error with product equip with a Dose Area Product (DAP) meter. This results in the product displaying the incorrect DAP meter values which may impede radiation exposure management decisions

Where Was This Sold?

This product was distributed nationwide across all 50 states.