Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Microbiologics QC Sets and Panels Recalled by Microbiologics Inc Due to The recommended hydration fluid for the 8180 Gram-Positive...

Date: December 16, 2019
Company: Microbiologics Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Microbiologics Inc directly.

Affected Products

Microbiologics QC Sets and Panels, Gram-Positive Blood Culture Control Panel (Inactivated Pellet), Catalog # 8180 Used in test systems to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation.

Quantity: 810

Why Was This Recalled?

The recommended hydration fluid for the 8180 Gram-Positive Blood Culture Control Panel (Inactivated Pellet), has changed from sterile 0.85% saline or sterile blood culture media to sterile DI water or pH 7.2 phosphate buffer.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected