Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Microbiologics Enteric Pathogens Control Panel (Inactivated Pellet) Recalled by Microbiologics Inc Due to The negative control was contaminated with one of...

Date: January 8, 2024
Company: Microbiologics Inc
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Microbiologics Inc directly.

Affected Products

Microbiologics Enteric Pathogens Control Panel (Inactivated Pellet), REF 8184.

Quantity: 580 packages

Why Was This Recalled?

The negative control was contaminated with one of the positive control pathogens (Norovirus). This would result in users getting a positive call for Norovirus when running the negative control.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Microbiologics Inc

Microbiologics Inc has 66 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report