Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

MicroAire PAL¿ Cannula Product Usage - Power Aspiration Device Recalled by MicroAire Surgical Instruments, LLC Due to MicroAire has initiated a recall on product PAL-R4011XL...

Date: February 8, 2016
Company: MicroAire Surgical Instruments, LLC
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact MicroAire Surgical Instruments, LLC directly.

Affected Products

MicroAire PAL¿ Cannula Product Usage - Power Aspiration Device

Quantity: 145 manufactured and distributed

Why Was This Recalled?

MicroAire has initiated a recall on product PAL-R4011XL (40cm x 4mm Del Vecchio French Rasp cannula), due to our receipt of three (3) complaints that it bent or broke during surgery. Accordingly, the PAL-R4011XL has also been put on production hold until further notice.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About MicroAire Surgical Instruments, LLC

MicroAire Surgical Instruments, LLC has 14 total recalls tracked by RecallDetector.

Related Recalls