Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
MEVION S250i Recalled by Mevion Medical Systems, Inc. Due to Couch Correction moves sent after using a 3D...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact Mevion Medical Systems, Inc. directly.
Affected Products
MEVION S250i, MEVION S250; Proton Radiation Treatment System
Quantity: 8 units
Why Was This Recalled?
Couch Correction moves sent after using a 3D CT scan are partially lost if the previous correction sent or recorded had some of the Couch Correction checkboxes turned off
Where Was This Sold?
This product was distributed to 6 states: FL, MO, NJ, OH, OK, DC
About Mevion Medical Systems, Inc.
Mevion Medical Systems, Inc. has 14 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report