Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

MEVION S250 Radiation therapy system Recalled by Mevion Medical Systems, Inc. Due to The treatment table positioning belt can slip gear...

Date: November 18, 2015
Company: Mevion Medical Systems, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Mevion Medical Systems, Inc. directly.

Affected Products

MEVION S250 Radiation therapy system

Quantity: 4

Why Was This Recalled?

The treatment table positioning belt can slip gear positions, resulting in deviation of patient setup and treatment positions from the time of slippage until the discovery of slippage and the halting of treatment.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected