Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

MEVION S250 Proton Radiation Treatment System Recalled by Mevion Medical Systems, Inc. Due to The perforated screen that divides the inside of...

Date: June 23, 2020
Company: Mevion Medical Systems, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Mevion Medical Systems, Inc. directly.

Affected Products

MEVION S250 Proton Radiation Treatment System; intended to deliver proton radiation treatment to patients

Quantity: 8 units

Why Was This Recalled?

The perforated screen that divides the inside of the process water tank has a very sharp edge across the top of the Heat Exchanger leading to potential injury for servicer.

Where Was This Sold?

This product was distributed to 6 states: FL, MO, NJ, OH, OK, DC

Affected (6 states)Not affected

About Mevion Medical Systems, Inc.