Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
Pentron Clinical Metrex Research Recalled by Metrex Research, LLC. Due to The firm initiated the recall due to a...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact Metrex Research, LLC. directly.
Affected Products
Pentron Clinical Metrex Research; Pinnacle Cure Sleeve, a protective barrier The intended use of these devices is to protect both patient and healthcare worker from transfer of microorganisms, body fluids and particulate matter.
Quantity: 3,348 units
Why Was This Recalled?
The firm initiated the recall due to a labeling error. The boxes labeled as Pinnacle Cure Sleeve 4511, which fits 11mm light guides, actually contain Pinnacle Cure Sleeves that fit 7mm or 8mm light guides (part number 4500).
Where Was This Sold?
This product was distributed nationwide across all 50 states.
About Metrex Research, LLC.
Metrex Research, LLC. has 3 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report