Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Mesa Smart-Read EZTEST BIOLOGICAL INDICATOR MONITORING SYSTEM. Intended for use Recalled by Mesa Laboratories, Inc., Bozeman Manufacturing Facility Due to Self-contained steam biological indicator, are recalled because Mesa...

Date: June 4, 2013
Company: Mesa Laboratories, Inc., Bozeman Manufacturing Facility
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Mesa Laboratories, Inc., Bozeman Manufacturing Facility directly.

Affected Products

Mesa Smart-Read EZTEST BIOLOGICAL INDICATOR MONITORING SYSTEM. Intended for use in determining the efficacy of steam sterilization processes.

Quantity: 2447/100 count boxes

Why Was This Recalled?

Self-contained steam biological indicator, are recalled because Mesa recently determined that the read out time is not in specification with the label claims.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Mesa Laboratories, Inc., Bozeman Manufacturing Facility

Mesa Laboratories, Inc., Bozeman Manufacturing Facility has 2 total recalls tracked by RecallDetector.

Related Recalls