Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Merit Medical Cultura Collection and Transport System Recalled by Merit Medical Systems, Inc. Due to The flocked tip of the 100mm swab may...

Date: September 16, 2020
Company: Merit Medical Systems, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Merit Medical Systems, Inc. directly.

Affected Products

Merit Medical Cultura Collection and Transport System, Ref CFTS100, Single Use, IVD, for the following unit counts: 1 Unit - (01)00884450492123, 50 Units - (01)10884450492120, 200 units - (01)20884450492127

Quantity: 159,396 units

Why Was This Recalled?

The flocked tip of the 100mm swab may break off in the nasophyaryngeal cavity during sample collection. This could result in soft tissue injury of the nasophyaryngeal cavity or a delay in treatment.

Where Was This Sold?

This product was distributed nationwide across all 50 states.