Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Merit Maestro Microcatheter Recalled by Merit Medical Systems, Inc. Due to These microcatheters were packaged with a 2.8 F...

Date: September 26, 2014
Company: Merit Medical Systems, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Merit Medical Systems, Inc. directly.

Affected Products

Merit Maestro Microcatheter, 2.4F 130cm Straight, REF28MC24130ST, Rx Only, Non-pyrogenic Sterile if package is unopened or undamaged.

Quantity: 56 units

Why Was This Recalled?

These microcatheters were packaged with a 2.8 F distal tip, rather than the labeled 2.4 F.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Merit Medical Systems, Inc.

Merit Medical Systems, Inc. has 135 total recalls tracked by RecallDetector.

Related Recalls