Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

The Schiller PB 1000 Recalled by Merge Healthcare, Inc. Due to Non-invasive blood pressure pumps (NIBP) voltage is not...

Date: May 31, 2015
Company: Merge Healthcare, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Merge Healthcare, Inc. directly.

Affected Products

The Schiller PB 1000, also known as: Merge Hemo Patient Data Module (PDM), V2 model, a major component of the Merge Hemo software.

Quantity: 341 sites potentially have the affected product versions

Why Was This Recalled?

Non-invasive blood pressure pumps (NIBP) voltage is not sufficient and causes the NIBP pumps to fail prior to reaching adequate initial pressure.

Where Was This Sold?

This product was distributed nationwide across all 50 states.