Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Schiller AG ISA CO2 Sensor (AKA: PhaseIn Unit) Recalled by Merge Healthcare, Inc. Due to The units may display the alarm "sampling line...

Date: March 30, 2015
Company: Merge Healthcare, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Merge Healthcare, Inc. directly.

Affected Products

Schiller AG ISA CO2 Sensor (AKA: PhaseIn Unit), Merge Part Number HW-HEMO-00001. The PhaseIn Unit is an optional accessory (sold separately) to the Merge Hemo Patient Data Module (PDM). The PDM is a major component of the Merge Hemo system.

Quantity: 23 sensors (PhaseIn Units)

Why Was This Recalled?

The units may display the alarm "sampling line clogged" after the flow zeroing procedure that is conducted automatically 1 hour after start up. This could result in the potential of a small bias added to the CO2 measurement and affect the alarm notification.

Where Was This Sold?

This product was distributed to 10 states: AL, FL, ID, LA, MO, NC, OK, SC, TX, UT