Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
Merge OrthoPACS software. Recalled by Merge Healthcare, Inc. Due to A migrated study that has annotations will display...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact Merge Healthcare, Inc. directly.
Affected Products
Merge OrthoPACS software.
Quantity: 98 sites potentially have the affected versions
Why Was This Recalled?
A migrated study that has annotations will display the annotations, but when a prior study is loaded that should have annotations, the second study will not display annotations.
Where Was This Sold?
This product was distributed nationwide across all 50 states.
About Merge Healthcare, Inc.
Merge Healthcare, Inc. has 93 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report