Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Merge Eye Station and Merge Eye Care PACS. Recalled by Merge Healthcare, Inc. Due to During an antivirus program scan of the Eye...

Date: April 4, 2016
Company: Merge Healthcare, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Merge Healthcare, Inc. directly.

Affected Products

Merge Eye Station and Merge Eye Care PACS.

Quantity: 1,627 sites potentially have the affected versions

Why Was This Recalled?

During an antivirus program scan of the Eye Station or Eye Care PACS, the antivirus program may detect a newly captured image to be an unwanted file and delete the image permanently.

Where Was This Sold?

This product was distributed nationwide across all 50 states.