Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

MegaPower Electrosurgical Generator Recalled by Megadyne Medical Products, Inc. Due to If two devices are plugged into the same...

Date: May 15, 2018
Company: Megadyne Medical Products, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Megadyne Medical Products, Inc. directly.

Affected Products

MegaPower Electrosurgical Generator, Product code 1000, GTIN 10614559103715

Quantity: 8380 units

Why Was This Recalled?

If two devices are plugged into the same channel/port in the generator, both devices activate. If one device is left laying on the patient, the result could be burning of the patient.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Megadyne Medical Products, Inc.

Megadyne Medical Products, Inc. has 20 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report