Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
Endurant / Endurant II 23mm and 25mm Bifurcated Stent Graft Recalled by Medtronic Vascular, Inc. Due to Medtronic is initiating a voluntary Urgent Medical Device...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact Medtronic Vascular, Inc. directly.
Affected Products
Endurant / Endurant II 23mm and 25mm Bifurcated Stent Graft Systems Model Numbers: ETBF2516C145E ETBF2516C166E ETBF2316C145E ETBF2313C145E ETBF2513C166E ETBF2313C166E ETBF2316C166E ETBF2313C124E ETBF2513C145E ETBF2516C124E ETBF2316C124E ETBF2513C124E. Cardiology: The Endurant family of stent graft systems is Intended to treat infra-renal abdominal aortic or aortoiliac aneurysms using an endovascular approach. When placed within the aneurysm, the stent graft provides an alternative conduit for blood flow within the patient's vasculature.
Quantity: 3331 devices
Why Was This Recalled?
Medtronic is initiating a voluntary Urgent Medical Device Recall for a subset of Endurant/ Endurant II Bifurcated Stent Graft Systems of specific models and serial numbers as this specific subset of stent grafts has greater susceptibility to fabric permeability variations that may be associated with endoleaks observed during the initial implant procedure.
Where Was This Sold?
Worldwide Distribution - US Distribution including Puerto Rico and to the countries of : Argentina Armenia Australia Austria Belgium Brazil Bulgaria Canada China Colombia Costa Rica Croatia Cyprus Czech Republic Denmark Dominican Republic Ecuador Egypt El Salvador Finland France French Polynesia Germany Greece Hong Kong Hungary Iceland India Indonesia Iran Ireland Israel Italy Japan Jordan Kazakhstan Korea Kuwait Latvia Lebanon Lithuania Luxembourg Malaysia Mexico Montenegro Netherlands New Zealand Norway Oman Panama Paraguay Peru Philippines Poland Portugal Reunion Romania Russian Federation Saudi Arabia Serbia Singapore Slovakia Slovenia South Africa Spain Sri Lanka Sweden Switzerland Taiwan Thailand Turkey United Arab Emirates United Kingdom Uruguay Venezuela Viet Nam
About Medtronic Vascular, Inc.
Medtronic Vascular, Inc. has 19 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report