Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Catalyft PL & PL40 Expandable Interbody System labeled as follows Recalled by Medtronic Sofamor Danek USA Inc Due to Expandable interbody system cage may loose height or...

Date: September 16, 2025
Company: Medtronic Sofamor Danek USA Inc
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Medtronic Sofamor Danek USA Inc directly.

Affected Products

Catalyft PL & PL40 Expandable Interbody System labeled as follows with corresponding model numbers: 1. SPACER 6068073 CATALYFT PL SHORT 7MM. Model Number: 6068073. 2. SPACER 6068076 CATALYFT PL LONG 7MM, Model Number: 6068076. 3. SPACER 6068093 CATALYFT PL SHORT 9MM, Model Number: 6068093. 4. SPACER 6068096 CATALYFT PL LONG 9MM, Model Number: 6068096. 5. SPACER 6068113 CATALYFT PL SHORT 11MM, Model Number: 6068113. 6. SPACER 6068116 CATALYFT PL LONG 11MM, Model Number: 6068116. 7. SPACER 6069073 CATALYFT PL40 SHORT 7MM, Model Number: 6069073. 8. SPACER 6069076 CATALYFT PL40 LONG 7MM, Model Number: 6069076. 9. SPACER 6069093 CATALYFT PL40 SHORT 9MM, Model Number: 6069093. 10. SPACER 6069096 CATALYFT PL40 LONG 9MM, Model Number: 6069096. 11. SPACER 6069113 CATALYFT PL40 SHORT 11MM, Model Number: 6069113. 12. SPACER 6069116 CATALYFT PL40 LONG 11MM, Model Number: 6069116. 13. CATALYFT PL IMPLANT SET, Model Number: SPS03143. 14. CATALYFT PL 40 IMPLANT SET, Model Number: SPS03144. 15. KIT CATALYFT PL IMPLANTS, Model Number: SPS04012. 16. KIT CATALYFT PL 40 IMPLANTS, Model Number: SPS04013. 17. CATALYFT PL 40 IMP KIT, Model Number: SPSJEP40. 18. CATALYFT PL 40 HIGH IMP KIT, Model Number: SPSJEP4H. 19. CATALYFT PL 40 SHORT IMP KIT, Model Number: SPSJEP4S. 20. CATALYFT PL 40 SHORT HIGH IMP KIT, Model Number: SPSJEP4T. 21. CATALYFT PL STRAIGHT HIGH IMP KIT, Model Number: SPSJEPSH. 22. CATALYFT PL STRAIGHT IMP KIT, Model Number: SPSJEPSI. These systems utilize the following affected Instructions for Use (IFU), Document M333023W048E, and Surgical Technique Guide, Document M333023W217.

Quantity: 275,260 units

Why Was This Recalled?

Expandable interbody system cage may loose height or collapse (loss of lordosis), which could result in migration, subsidence and/or neurological injury. Therefore, surgical technique and instructions for use will be updated to mitigate risk of loss of lordosis to occur.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Medtronic Sofamor Danek USA Inc

Medtronic Sofamor Danek USA Inc has 81 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report