Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Braive Growth Modulation System Break-Off Set Screws - Product Usage: Recalled by Medtronic Sofamor Danek USA, Inc Due to Product is impacted by a thread profile defect...

Date: April 27, 2021
Company: Medtronic Sofamor Danek USA, Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Medtronic Sofamor Danek USA, Inc directly.

Affected Products

Braive Growth Modulation System Break-Off Set Screws - Product Usage: intended to limit further progression of the spinal curvature and provide correction of the thoracic curve while allowing continued skeletal growth by compression of the convex side of the scoliotic curve.

Quantity: 162 UNITS

Why Was This Recalled?

Product is impacted by a thread profile defect due to a manufacturing issue.

Where Was This Sold?

International distribution in the countries of France, Switzerland and Jordan. No US distribution.

About Medtronic Sofamor Danek USA, Inc

Medtronic Sofamor Danek USA, Inc has 4 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report