Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Custom Pack Perfusion Tubing Packs containing Affinity Fusion Oxygenator with Recalled by Medtronic Perfusion Systems Due to An upward trend of complaints in which the...

Date: December 12, 2022
Company: Medtronic Perfusion Systems
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Medtronic Perfusion Systems directly.

Affected Products

Custom Pack Perfusion Tubing Packs containing Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface, REF CB811. Used in Cardiopulmonary bypass.

Quantity: 40,498 units

Why Was This Recalled?

An upward trend of complaints in which the Temperature Monitoring Adapter (TMA), located on the Affinity Fusion Oxygenator, had come loose from the oxygenator either during pre-procedure perfusion set-up or post-procedure when disassembling the perfusion circuit.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected