Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Model B35200 - Percept PC BrainSense Implantable Neurostimulator Recalled by Medtronic Neuromodulation Due to During cardioversion events, the Model B35200 Percept PC...

Date: October 19, 2021
Company: Medtronic Neuromodulation
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Medtronic Neuromodulation directly.

Affected Products

Model B35200 - Percept PC BrainSense Implantable Neurostimulator

Quantity: 10,006 units

Why Was This Recalled?

During cardioversion events, the Model B35200 Percept PC Implantable Neurostimulator could become unresponsive and non-functional. Cardioversion is a procedure performed by a cardiologist to correct an abnormal heartbeat. Medtronic has determined that the cardioversion may damage the circuit electronics in the Percept PC stimulator, thereby, causing the Percept PC stimulator to become non-responsive.

Where Was This Sold?

This product was distributed nationwide across all 50 states.