Class I — Serious Health Hazard

Serious health hazard — there is a reasonable probability that use of or exposure to this product will cause serious adverse health consequences or death.

Medtronic Sutureless Pump Connector Revision Kit Recalled by Medtronic Neuromodulation Due to The Sutureless Connector (SC) Intrathecal Catheter connector has...

Date: June 3, 2013
Company: Medtronic Neuromodulation
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Medtronic Neuromodulation directly.

Affected Products

Medtronic Sutureless Pump Connector Revision Kit, model 8578. Contents: catheter interface with attached sutureless pump connector, catheter, connector pin, and strain-relief sleeve to be used with Medtronic SynchroMed implantable drug infusion pumps. Contents of inner package are STERILE. The Medtronic Model 8578 Sutureless Pump Connector Revision Kit is used when a pump connector for an Indura 1P Model 8709 or Model 8709SC catheter is required. The catheter is part of an infusion system that stores and delivers parenteral drugs to the intrathecal space. The implanted infusion system components consist of a Medtronic pump and an Indura 1P Model 8709 or Model 8709SC catheter. The catheter connects to the pump with the Model 8578 sutureless pump connector at the catheter port.

Quantity: 115,722 total

Why Was This Recalled?

The Sutureless Connector (SC) Intrathecal Catheter connector has been redesigned to reduce the potential for occlusion at the catheter to pump interface. Medtronic is removing the unused products from the market that were manufactured with the previous design, and recommend the previous design no longer be used due to greater potential for misalignment and subsequent occlusion.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Medtronic Neuromodulation

Medtronic Neuromodulation has 95 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report