Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Medtronic Recalled by Medtronic Neuromodulation Due to Medtronic neurostimulators (used with pain and deep brain...

Date: February 27, 2014
Company: Medtronic Neuromodulation
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Medtronic Neuromodulation directly.

Affected Products

Medtronic, External neurostimulator, ENS (37021) & ENS (37022). Including the external neurostimulators which are associated with software application run on the NVision Clinician Programmer and used during trialing for spinal cord stimulation (SCS) patients.

Quantity: 170,286 US, 73,921 OUS for all devices.

Why Was This Recalled?

Medtronic neurostimulators (used with pain and deep brain stimulation therapy) have a cycling feature that, when enabled, allows stimulation output to be toggled on and off based on the programmed settings. The use of this feature is determined by the physician and set using the clinician programmer. Current labeling indicates the use of cycling improves device longevity and recharge interval

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected