Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
A610 Clinician Programmer Application for Deep Brain Stimulation Clinician Programming Recalled by Medtronic Neuromodulation Due to There are two issues within this correction action:...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact Medtronic Neuromodulation directly.
Affected Products
A610 Clinician Programmer Application for Deep Brain Stimulation Clinician Programming Guide Percept PC Model B35200 Neurostimulator Application, version 2.0.4584, 2.0.4594, 2.0.4605, and 3.0.1057* * Version 3.0.1057 is approved for distribution in Europe only.
Quantity: 5332 downloads
Why Was This Recalled?
There are two issues within this correction action: 1. Cycling Issue: When copying settings from a previously used Activa device to a new Percept PC Implantable Neurostimulator (INS) using the "replacement" function of the A610 clinician programmer application, the cycling feature is not copied properly. 2. Log Fill Issue: When the Percept PC INS System Event log is full, the clinician programmer A610 Software application version 2.0.4584, 2.0.4594, 2.0.4605, or 3.0.1057 crashes when it attempts a reading of the last records in the log.
Where Was This Sold?
This product was distributed nationwide across all 50 states.
About Medtronic Neuromodulation
Medtronic Neuromodulation has 95 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report