Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Medtronic 0-arm 02 Imaging System Catalog Number: Bl-700-02000 The O-arm Recalled by Medtronic Navigation, Inc. Due to Screws holding the detector panel in as a...

Date: January 29, 2016
Company: Medtronic Navigation, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Medtronic Navigation, Inc. directly.

Affected Products

Medtronic 0-arm 02 Imaging System Catalog Number: Bl-700-02000 The O-arm O2 Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging and is intended to be used where a physician benefits from 2D and 3D information of anatomic structures and objects with high x-ray attenuation such as bony anatomy and metallic objects. The O-arm O2 Imaging System is compatible with certain image guided surgery systems.

Quantity: 1 unit

Why Was This Recalled?

Screws holding the detector panel in as a result the detector panel could make contact with other parts with in the 0-arm¿ 02 Imaging System, causing noise and vibration when being positioned or during 3D Imaging could lead to poor image quality

Where Was This Sold?

Worldwide Distribution to Austria only.

About Medtronic Navigation, Inc.

Medtronic Navigation, Inc. has 91 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report