Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
MiniMed 780G Insulin Pump: Models: PUMP MMT-1885 MM780G 6.5V BLE MMOL Recalled by Medtronic Minimed Due to Due to a software design issue, under certain...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact Medtronic Minimed directly.
Affected Products
MiniMed 780G Insulin Pump: Models: PUMP MMT-1885 MM780G 6.5V BLE MMOL;¿ PUMP MMT-1886 MM780G 6.5W BLE MG; KIT MMT-1895WW MM780G 6.5V BLE SF MMOL;¿ KIT MMT-1896WW MM780G 6.5W BLE SF MG;¿ PUMP MMT-1884XC 780G V6.5 CLIN US MG;¿ PUMP MMT-1885XC 780G V6.5 CLIN MM;¿ PUMP MMT-1886XC 780G V6.5 CLIN MG;¿ PUMP MMT-1885L MM780G 6.5V BLE MMOL;¿ PUMP MMT-1886L MM780G 6.5W BLE MG;¿ KIT MMT-1896ES MM780G V6.5W MG;¿ KIT MMT-1896WWA MM780G BLE MG ES;¿ KIT MMT-1895WWA MM780G BLE MMOL AR;¿ KIT MMT-1895WWA MM780G BLE MMOL CS;¿ KIT MMT-1895WWA MM780G BLE MMOL DA;¿ KIT MMT-1895WWA MM780G BLE MMOL DE;¿ KIT MMT-1895WWA MM780G BLE MMOL EN;¿ KIT MMT-1895WWA MM780G BLE MMOL FI;¿ KIT MMT-1895WWA MM780G BLE MMOL FR;¿ KIT MMT-1895WWA MM780G BLE MMOL IT;¿ KIT MMT-1895WWA MM780G BLE MMOL NL;¿ KIT MMT-1895WWA MM780G BLE MMOL NO;¿ KIT MMT-1895WWA MM780G BLE MMOL SL; KIT MMT-1895WWA MM780G BLE MMOL SV;¿ KIT MMT-1895WWA MM780G BLE MMOL HU;¿ KIT MMT-1896WWA MM780G BLE MG AR;¿ KIT MMT-1896WWA MM780G BLE MG DE;¿ KIT MMT-1896WWA MM780G BLE MG EL;¿ KIT MMT-1896WWA MM780G BLE MG HE;¿ KIT MMT-1896WWA MM780G BLE MG IT;¿ KIT MMT-1896WWA MM780G BLE MG PL;¿ KIT MMT-1896WWA MM780G BLE MG RO;¿ KIT MMT-1896WWA MM780G BLE MG TR;¿ KIT MMT-1896WWA MM780G BLE MG NL;¿ KIT MMT-1896WWA MM780G BLE MG FR;¿ KIT MMT-1896WWA MM780G BLE MG ES;
Quantity: O.U.S. Distibution only = 15,682 devices
Why Was This Recalled?
Due to a software design issue, under certain conditions, a software fault is detected when a large bolus delivery at quick bolus speed completes. if the user is not aware of the amount of active insulin and delivers an additional bolus, there is a risk of insulin over delivery.
Where Was This Sold?
This product was distributed to 12 states: AR, CA, CO, CT, FL, GA, ID, MI, MN, TN, TX, WA
About Medtronic Minimed
Medtronic Minimed has 2 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report