Class I — Serious Health Hazard

Serious health hazard — there is a reasonable probability that use of or exposure to this product will cause serious adverse health consequences or death.

MiniMed 630G Insulin Pump Recalled by Medtronic MiniMed, Inc. Due to Insulin pumps that have been dropped, bumped, or...

Date: July 31, 2024
Company: Medtronic MiniMed, Inc.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Medtronic MiniMed, Inc. directly.

Affected Products

MiniMed 630G Insulin Pump, REF: MMT-1515, MMT-1714, MMT-1715, MMT-1754, MMT-1755; MiniMed 700G Insulin Pump, REF: MMT-1800, MMT-1801, MMT-1805, MMT-1850, MMT-1851

Quantity: 24,595

Why Was This Recalled?

Insulin pumps that have been dropped, bumped, or experienced physical impact may have damage to internal electrical components, which may cause reduced pump battery life. Since the defect reduces the battery life overall, it shortens the time between low battery alarms and increases the frequency with which the user must insert a new battery. Further, the defect may also deliver the low battery alarm when there is significantly less time before the battery completely runs out, from the expected up to 10-hour buffer the low alert is supposed to signal, leading to device powering down and stopping insulin delivery unexpectedly. Per the firm, a subset of pumps with the defect showed the low alert alarm was delivered ~2.5 hours before the device shut off, however, patient reports indicate the time could be even shorter. Replacing the battery will not resolve this issue and the short battery life and the truncated low battery alert lead time will continue to occur. Early battery depletion could result in power loss and insulin delivery could be interrupted, resulting in under delivery of insulin potentially leading to hyperglycemia and/or DKA.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Medtronic MiniMed, Inc.

Medtronic MiniMed, Inc. has 22 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report