Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Medtronic REF: MMT-7020C5 Guardian Sensor (3) for the following packaging: Recalled by Medtronic Inc. Due to Use of the continuous glucose monitoring system while...

Date: February 26, 2021
Company: Medtronic Inc.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Medtronic Inc. directly.

Affected Products

Medtronic REF: MMT-7020C5 Guardian Sensor (3) for the following packaging: GS3 5PK 1L AUS/ GTIN: 20763000252056 GS3 5PK 1L AUS/ GTIN: 20763000414362 Product Usage: intended for use with Medtronic Diabetes (Medtronic) glucose sensing systems to continuously monitor glucose levels in persons with diabetes. Product Usage: intended for use with Medtronic Diabetes (Medtronic) glucose sensing systems to continuously monitor glucose levels in persons with diabetes. *Product not distributed within US.

Quantity: 68,069 units

Why Was This Recalled?

Use of the continuous glucose monitoring system while actively taking hydroxyurea, an anti-neoplastic drug could result in inaccurate elevated sensor glucose readings.

Where Was This Sold?

Worldwide distribution.