Medtronic CareLink iPro Therapy Management Software Recalled by Medtronic Inc. Due to Due to a time conversion error, data uploaded...

Date: October 10, 2017
Company: Medtronic Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Medtronic Inc. directly.

Affected Products

Medtronic CareLink iPro Therapy Management Software, MMT-7340 The CareLink iPro Therapy Management Software (model, MMT-7340) is intended to be used by a healthcare professional (HCP) with the iPro2 CGM System and provides upload support for the iPro2 Recorder. The CareLink iPro software is used to upload glucose sensor data stored on a recording device, capture BG meter readings through manual entry or wired-upload, retrospectively calibrate sensor data, and provide retrospective reports of continuous glucose information. The reports generated by the CareLink iPro software are intended for use by healthcare professionals to assist with diabetes clinical assessment, patient education, and therapy optimization.

Quantity: 488 uploads

Why Was This Recalled?

Due to a time conversion error, data uploaded from the iPro(TM)2 recorder to the CareLink iPro software using the new CareLink iPro Uploader feature may result in incorrect reports. This pertains only to reports generated between September 23, 2017 and September 27, 2017 that used the new CareLink iPro Uploader feature.