Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Guardian Connect App CSS7200 used on the iPhone Recalled by Medtronic Inc. Due to The application may be closed by the operating...

Date: July 4, 2018
Company: Medtronic Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Medtronic Inc. directly.

Affected Products

Guardian Connect App CSS7200 used on the iPhone, iPad, and iPod Touch devices.

Quantity: Approximately 9,328 registered users

Why Was This Recalled?

The application may be closed by the operating system without alerting the user the app is no longer running or communicating with the transmitter resulting in the user not receiving alerts that could be associated with hypoglycemic or hyperglycemic events.

Where Was This Sold?

Distribution was only made OUS. Countries involved are: Austria, United Arab Emirate, Australia, Belgium, Switzerland, Chile, Colombia, Czech Republic, Germany, Denmark, Spain, Finland, France, United Kingdom, Hungary, Ireland, Israel, Iceland, Italy, Republic of Korea, Luxembourg, Netherlands, Norway, Poland, Qatar, Romania, Russian Federation, Saudi Arabia, Sweden, Singapore, Slovenia, Slovakia, Turkey, and South Africa.