Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Medtronic Power Supply 26907 Recalled by Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) Due to There is a potential for an electrical short...

Date: January 17, 2019
Company: Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) directly.

Affected Products

Medtronic Power Supply 26907, Medtronic CareLink Encore(TM) 29901 Programmer

Quantity: 731 units

Why Was This Recalled?

There is a potential for an electrical short circuit to develop in a subset of 479 Model 26907 power supply cable plugs that connect to the Medtronic Model 29901 CareLink Encore programmer. A failure mode of the power supply has been found where an electrical short occurs on the low voltage direct current connector and results in the connector heating up, generating smoke, melting and in some cases disconnecting from the programmer.

Where Was This Sold?

US and Armenia, Bulgaria, Croatia, Georgia, Hungary, India, Iran, Jordan, Kazakhstan, Latvia, Lebanon, Oman, Pakistan, Qatar, Romania, Saudi Arabia, Syrian Arab Republic

About Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) has 198 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report