Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Azure S SR MRI Recalled by Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) Due to There is a rare communication sequence during the...

Date: June 4, 2020
Company: Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) directly.

Affected Products

Azure S SR MRI, Model Number W3SR01

Quantity: 11953 devices

Why Was This Recalled?

There is a rare communication sequence during the first device interrogation with a SmartSync Device Manager that may result in the temporary suspension of some device features (i.e., battery measurements, Capture Management", Atrial Lead Position Check", EffectivCRT" algorithms, and AdaptivCRT"). This rare interaction results in temporary suspension of automatic threshold testing and output adjustments, and suspension of auto-optimization of CRT therapy. The issue is unlikely to result in clinical impact to the patient, and features are restored upon next programmer device interrogation or presence of a magnet.

Where Was This Sold?

worldwide

About Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) has 198 total recalls tracked by RecallDetector.