Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

PROFILE-V MEDTOXScan Drugs of Abuse Test System. For the rapid Recalled by Medtox Diagnostics Inc Due to Review of validation information and customer complaints which...

Date: April 5, 2013
Company: Medtox Diagnostics Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Medtox Diagnostics Inc directly.

Affected Products

PROFILE-V MEDTOXScan Drugs of Abuse Test System. For the rapid, qualitative detection of drugs of abuse in human urine,

Quantity: 221,625

Why Was This Recalled?

Review of validation information and customer complaints which indicated preliminary screen positive results for cocaine that were not positive on confirmatory testing.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Medtox Diagnostics Inc

Medtox Diagnostics Inc has 1 total recall tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report